The customer, a syncardia authorized distributor, reported that the patient was performing routine morning activity under supervision of the nurses in the clinic when the freedom driver exhibited a continuous fault alarm.Clinic staff checked the freedom drivelines, connections and the patient's blood pressure but found nothing wrong.After the inspection, the freedom driver suddenly stopped working and was recognized immediately and the patient was switched to the backup driver.The customer also reported that the patient briefly lost consciousness but recovered quickly after the driver switch and there was no permanent health impact.
|
The customer-reported fault alarm followed by a driver stop was not able to be confirmed or reproduced during investigation testing.The driver's alarm history was reviewed and revealed one new alarm fault code.This 2d fault code was likely produced at syncardia during the last service of the driver as there was no evidence of secondary motor engagement.Only permanent fault alarms are recorded in the driver's alarm history.Intermittent and/or recoverable alarms are not recorded.The driver in "as received" condition passed all sections of functional testing.Additionally, an extended, seven day observation run was performed on the driver where it performed as intended with no abnormalities, alarms, interruptions in operation, or burning smells.The onboard batteries returned with the driver and associated with being in use by the patient were each individually evaluated and each passed all sections of the freedom onboard battery evaluation procedure.The driver and onboard batteries performed as intended with no evidence of a device malfunction.The root cause of the customer-reported issue could not be determined.This issue will continue to be monitored and trended as part of the customer experience process.Syncardia has completed its evaluation of this complaint and is closing this file.Ce 4880 follow-up report 1.
|