The customer contacted a siemens customer care center (ccc) and reported that the platelet (plt) calibration factor was set to 1.19 on an advia 2120 hematology system with dual aspirate autosampler (serial number: (b)(4)) in may 2019 in order for quality controls (qc) to recover within range.This is a use error as the calibration factor, although within range, should not be changed after being initially set in order to alter qc results.The system was put out of use once the calibration factor had been deemed incorrect and a siemens customer service engineer (cse) was dispatched to the customer site to investigate why qc were not recovering within range.The cse observed that the plt count results on the affected advia 2120 hematology system with dual aspirate autosampler (serial number: (b)(4)) were higher compared with an alternate instrument in the lab.The cse performed a periodic maintenance, replaced a leaking shear face and aligned it, replaced the unified fluidics chamber, replaced the red blood cell flowcell, verified the voltages on the signal processing board (spb) and replaced it.The cse tested the voltages on the existing preamp board which should be the same as the spb; however, the voltage was different when compared to the spb test points.The cse replaced the preamp board and then the voltages matched.The cse was then able to increase the preamp board voltages to -7.5v.The cse ran the original optipoint set gains, which were all within specification.The cse then adjusted the dispense and aspirate positions, after which plt counts decreased.The cause of the qc material to be out of range is unknown.Siemens is investigating the issue.
|
Siemens filed the initial mdr 2432235-2019-00253 on 01-aug-2019.Additional information (06-aug-2019): siemens further investigated the issue.The customer reported that there were no discordant patient results obtained as a result of this issue.Additionally, as per the online help advia 120/2120/2120i operator's guide v6.01 section calibration tab: calibration acceptance, an acceptable calibration factor must be between 0.50 and 1.50.In this case a platelet calibration factor was originally 1.19 and gave a higher platelet correlation result.The siemens customer service engineer performed optics bench adjustments and optical alignments, and these actions produced a platelet calibration factor of 1.08.The customer is fully operational.Result code and conclusion code in section h6 were updated based on the additional information.The system is performing according to specifications.No further evaluation of this device is required.
|