Model Number N/A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Unspecified Infection (1930); Necrosis (1971); Pain (1994); Swelling (2091)
Event Date 07/10/2019
Event Type  Injury  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 3007963827-2019-00217, 0001822565-2019-03230.Concomitant medical products: partial tibial cemented, item # 42538000801, lot # 64032073.Partial articular surface left medial, item # 42518200808, lot # 64090439.Customer has not indicated whether the product will be returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a patient was revised approximately two and a half months post implantation due to deep infection.There is no additional information at this time.
 
Manufacturer Narrative
This follow-up is being submitted to update the complaint description and patient codes.
 
Event Description
It was reported that a patient was revised approximately two and a half months¿ post implantation due to pain, swelling, necrosis, and deep infection of the left knee.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Updated: b4, b5, g4, g7, h1, h2, h3, h6, and h10.The reported event was confirmed through medical records.Primary surgery notes indicate no intra-operative complications.Three months¿ status post left medial uni-knee.X-rays report reveals well-positioned implants and no change from the six-week follow-up visit.Revision notes state that turbid fluid was seen that did communicate through a defect in the medial retinaculum of the knee joint.Implants were well fixed.All necrotic tissue was debrided and very little bone loss with the removal of the components.No devices were received; therefore the condition of the components is unknown.Device history record (dhr) was reviewed for deviations and/ or anomalies with no related deviations / anomalies identified.Root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PARTIAL FEMUR CEMENTED SIZE 6 LEFT MEDIAL
Type of Device
PROSTHESIS, KNEE
Manufacturer (Section D)
ZIMMER ORTHOPAEDIC MFG. LTD.
building no 2 east park
shannon industrial estate
shannon, county clare
EI 
MDR Report Key8861269
MDR Text Key153278350
Report Number3007963827-2019-00216
Device Sequence Number1
Product Code HSX
Combination Product (y/n)N
PMA/PMN Number
K161592
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup,Followup
Report Date 01/03/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number42558000601
Device Lot Number63944794
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/12/2019
Initial Date FDA Received08/05/2019
Supplement Dates Manufacturer Received08/01/2019
12/31/2019
Supplement Dates FDA Received08/27/2019
01/03/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Weight109
-
-