Model Number 60831 |
Device Problems
Off-Label Use (1494); Insufficient Flow or Under Infusion (2182)
|
Patient Problem
Tissue Damage (2104)
|
Event Date 08/01/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of a cardioblate bp2 device after making 2 to 3 lesions the operator noted that the saline could not reach the jaws and the target tissue burned.The operator checked and replaced the tubing but could not fix the issue.The device was replaced to complete the procedure.
|
|
Manufacturer Narrative
|
Visual inspection showed no outward signs of damage to the device.Additional visual inspection showed saline residue in the tubing.Performance analysis: a couple of ablation cycles were initiated.There was saline flow observed from the lower/fixed jaw.The device was cut apart and the upper jaw then started to operate.Conclusion: reason for return was confirmed for poor saline flow.This appears to be a one off issue.Review of complaint data for the 60831device has found no similar instances for this defect.There was no adverse effect to the patient.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information that during use of a cardioblate bp2 device, after making 2-3 lesions the operator noted that the saline could not reach the jaws and the target tissue burned.The operator checked and replaced the tubing but could not fix the issue.The device was replaced to complete the procedure.There was no adverse effect to the patient.
|
|
Search Alerts/Recalls
|