Model Number 429888 |
Device Problems
Device Contamination with Body Fluid (2317); Activation, Positioning or Separation Problem (2906)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 07/15/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
It was reported during implant of the left ventricular (lv) lead, the guidewire catheter could not pass through the lead due to suspected blood ingress.The lv lead was removed and replaced.No patient complications have been reported as a result of this event.
|
|
Manufacturer Narrative
|
Product event summary: the full lead was returned and analyzed.The analysis indicated that the distal end of the lead were extrinsically damaged.Visual analysis of the lead indicated damage at implant.The analyst noted the full lead was returned without the guidewire.Guidewire test could not be performed due to the distal tip seal damaged in the distal nose tip.If information is provided in the future, a supplemental report will be issued.
|
|
Search Alerts/Recalls
|