Model Number 71940-01 |
Device Problem
Appropriate Term/Code Not Available (3191)
|
Patient Problems
Abscess (1690); Erythema (1840); Unspecified Infection (1930); Itching Sensation (1943)
|
Event Date 07/17/2019 |
Event Type
Injury
|
Manufacturer Narrative
|
The product has been requested back for an investigation.A follow-up report will be submitted once additional information is obtained.The device mfg date is unknown.The date entered is the date abbott diabetes care became aware of the event.All pertinent information available to abbott diabetes care has been submitted.
|
|
Event Description
|
Customer reported an observing an abscess and experiencing symptoms of redness, itching, and "feels hot" at the site of the adc freestyle libre sensor, customer had contact with a healthcare professional to have the abscess "removed under general anesthesia" and was also provided ibuprofen.There was no report of death or permanent injury associated with this event.
|
|
Manufacturer Narrative
|
Since no product has been returned, extended investigation has been performed for the reported complaint and determined that there was no indication that the product did not meet specification.Dhrs (device history review) for the freestyle libre sensor and freestyle libre sensor kit were reviewed and the dhrs showed the freestyle libre sensor and freestyle libre sensor kit passed all tests prior to release.Dose audit reports were reviewed and demonstrates the continued effectiveness of the established sterilization process for libre sensor kits.Environmental monitoring reports were reviewed, including bioburden and endotoxin testing, and demonstrated that all monitoring processes continue to meet adc minimum requirements for product quality.All review activities conducted above, including but not limited to the final release testing specifically associated with the manufacture of this product, are sufficient information in order to show if the product has met specifications prior to release.If the product is returned, the case will be re-opened and a physical investigation will be performed.
|
|
Event Description
|
Customer reported an observing an abscess and experiencing symptoms of redness, itching, and "feels hot" at the site of the adc freestyle libre sensor, customer had contact with a healthcare professional to have the abscess "removed under general anesthesia" and was also provided ibuprofen.There was no report of death or permanent injury associated with this event.
|
|
Search Alerts/Recalls
|