Model Number CI-1601-05
Device Problems Product Quality Problem (1506); Defective Component (2292)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/18/2019
Event Type  malfunction  
Event Description
The recipient is reportedly experiencing poor sound quality due to open electrodes.Programming adjustments were made, however, the issue did not resolve.Revision surgery is scheduled.
 
Manufacturer Narrative
Advanced bionics considers the investigation into this reportable event as closed.The external visual inspection revealed the silicone and electrode were damaged at the fantail prior to receipt.The photographic imaging inspection revealed multiple damaged electrode wires at the fantail.System lock was verified.The electrode condition prevented some electrical tests from being performed.The device passed some of the electrical tests performed.The device passed the mechanical test performed.The scanning electron microscopy analysis confirmed multiple wires were cut and damaged at the fantail.The photographic imaging inspection and the scanning electron microscopy analysis revealed damaged silicone and electrode wires at the fantail region.It is believed that the damaged wires caused the open impedances measured in the operating room and at the clinic.Advanced bionics will continue to monitor failure modes of this type.This is the final report.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Manufacturer Narrative
The recipient's device was explanted.The recipient was reimplanted with another advanced bionics cochlear device.Disclaimer: advanced bionics does not intend that this report be any admission of liability, fault or product defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HIRES¿ ULTRA 3D IMPLANT
Type of Device
COCHLEAR IMPLANT
Manufacturer (Section D)
ADVANCED BIONICS, LLC
28515 westinghouse place
valencia CA 91355
MDR Report Key8875490
MDR Text Key153954330
Report Number3006556115-2019-00500
Device Sequence Number1
Product Code MCM
UDI-Device Identifier07630016862267
UDI-Public(01)07630016862267(11)190312(17)220228
Combination Product (y/n)N
PMA/PMN Number
P960058
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup,Followup
Report Date 07/18/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date02/28/2022
Device Model NumberCI-1601-05
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/16/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age69 YR
-
-