• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number C408646
Device Problem Material Twisted/Bent (2981)
Patient Problem Vascular Dissection (3160)
Event Date 07/29/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
After stent placement in the proximal left circumflex artery, a dissection occurred.During oct imaging for post stent assessment the catheter was attempted to be positioned, but it could not pass through the stent.The device was kinked after the monorail proximal of the stent in the vessel.The catheter was re-positioned but the issue persisted, so the device was removed from the patient¿s anatomy.The tip was inspected and found to be kinked.After removing the catheter, angio was performed to check the vessel and a dissection was discovered proximal of the first implanted stent where the device was kinked.A second stent was placed to cover the dissection.No optical coherence tomography was done.The patient is stable.
 
Manufacturer Narrative
One dragonfly optis imaging catheter was received for evaluation.The results of the investigation concluded that there were kinks between the lens and guidewire exit port, consistent with the reported event.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.The cause of the reported positioning issue is consistent with the kink.The cause of the kink is consistent with damage during use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key8876594
MDR Text Key153789256
Report Number3008452825-2019-00397
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (y/n)N
PMA/PMN Number
K141769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/08/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model NumberC408646
Device Lot Number6964433
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received08/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-