• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 25AGFN-756
Device Problem Difficult to Insert (1316)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/16/2019
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2019, a 25 mm regent heart valve was selected for implant.During implant, the physician was unable to get the valve to rotate and exchanged the valve for another unknown valve.Due to the valve unable to rotate, there was a clinically significant delay in procedure.No patient consequences were reported.
 
Manufacturer Narrative
An event of inability to rotate the valve could not be confirmed.The investigation confirmed the device met functional specifications when analyzed at abbott.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.The cause of the reported event could not be conclusively determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key8878866
MDR Text Key153838110
Report Number2648612-2019-00058
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734006262
UDI-Public05414734006262
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/07/2020
Device Model Number25AGFN-756
Device Catalogue Number25AGFN-756
Device Lot Number5045748
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/30/2019
Was the Report Sent to FDA? No
Date Manufacturer Received09/15/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-