Udi -- (b)(4).One forcep was returned for evaluation with minor scissoring observed.On 07/30/2019, the forcep was visually inspected to check for tip alignment.The tips were slightly misaligned.Therefore, this complaint is confirmed.However, the root cause is unable to be determined.The complaint condition was replicated.The device did contribute to the complaint condition.The device did not meet specification.Device history records (dhrs) were reviewed and no anomalies were found.Based on the results of the investigation, the reported issue was confirmed.
|