Catalog Number 0684-00-0474 |
Device Problems
Defective Component (2292); Difficult to Advance (2920)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 07/16/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Complete event site name - (b)(6) university.The product has been returned to the manufacturer, but is pending investigation.Once the investigation is completed a supplemental report with our findings will be submitted.Complaint # (b)(4).
|
|
Event Description
|
It was reported that during insertion of the intra-aortic balloon (iab), an obstruction was found along the guidewire.The doctor checked the balloon and found the lumen was broken.The balloon was replaced to continue therapy.
|
|
Manufacturer Narrative
|
The product was returned with the membrane completely folded with traces of blood on the exterior & interior of the catheter.The one-way valve was also returned and still attached to the extracorporeal tubing.Two (2) inner lumen kinks were observed at approximately 7.6 & 13.7cm from the iab tip.A break in the inner lumen was also observed which appeared to have been the result of a severe kink at approximately 77.2cm from the iab tip.Lastly, a catheter kink was observed closest to the y-fitting at approximately 77.2cm from the iab tip.The one-way valve was vacuum tested and it held vacuum.An underwater leak test of the balloon, catheter, y-fitting, extracorporeal, extender and pressure tubing was performed and no leaks were detected.The break found in the inner lumen appears to have been the result of a severe kink, which eventually failed causing the reported difficulty to advance the iab through the guidewire.It is difficult to determine when the break occurred but it is possibly a result of patient movement during the procedure.A device and lot history record review was completed for the reported product.No non-conformances were found that are considered to be related to the event.Reference complaint #(b)(4).
|
|
Event Description
|
It was reported that during insertion of the intra-aortic balloon (iab), an obstruction was found along the guidewire.The doctor checked the balloon and found the lumen was broken.The balloon was replaced to continue therapy.
|
|
Search Alerts/Recalls
|