• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ROCHE DIAGNOSTICS ELECSYS FT4 III ASSAY; RADIOIMMUNOASSAY, FREE THYROXINE Back to Search Results
Model Number FT4 G3
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 07/18/2019
Event Type  malfunction  
Manufacturer Narrative
This event occurred in (b)(6).
 
Event Description
The initial reporter complained of questionable results for 46 patient samples tested for ft4 iii assay on a cobas 8000 e 801 module compared to the beckman dxl.Of the data provided, there were 7 patient samples with discrepant ft4 iii and elecsys tsh assay results, 21 patient samples with discrepant ft4 iii results, and 9 patient samples with discrepant tsh results.This medwatch will cover ft4 iii.Refer to medwatch with patient identifier (b)(6) for information on the tsh results.The e 801 module serial number was (b)(4).
 
Manufacturer Narrative
No samples were returned for investigation.Differences in results and reference ranges from ft4 assays by different manufactures, in this case abbott, can vary due to differences in the types of antibodies used, setups of the assays, and standardization methodologies.The investigation did not identify a product problem.The cause of the event could not be determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ELECSYS FT4 III ASSAY
Type of Device
RADIOIMMUNOASSAY, FREE THYROXINE
Manufacturer (Section D)
ROCHE DIAGNOSTICS
9115 hague road
indianapolis IN 46250 0457
MDR Report Key8887466
MDR Text Key154169558
Report Number1823260-2019-02962
Device Sequence Number1
Product Code CEC
Combination Product (y/n)N
PMA/PMN Number
K131244
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial,Followup
Report Date 01/07/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberFT4 G3
Device Catalogue Number07976887190
Device Lot NumberASKU
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 07/18/2019
Initial Date FDA Received08/13/2019
Supplement Dates Manufacturer Received07/18/2019
Supplement Dates FDA Received01/07/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-