• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ETHICON INC. INTERCEED* XL (TC7); BARRIER, ABSORBABLE, ADHESION

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ETHICON INC. INTERCEED* XL (TC7); BARRIER, ABSORBABLE, ADHESION Back to Search Results
Catalog Number 4350XL
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Abscess (1690); Unspecified Infection (1930); Inflammation (1932); Not Applicable (3189)
Event Date 07/15/2019
Event Type  Injury  
Manufacturer Narrative
Product complaint # (b)(4).To date the device has not been returned.If the device or further details are received at the later date a supplemental medwatch will be sent.Additional information was requested, and the following was obtained: were there any pre-existing signs/symptoms of active infection prior to this surgical procedure? no further information is available.Did the patient receive any prophylactic antibiotics pre-, intra- or post-operation? no further information is available.What was a cause of infection on 3rd day after initial surgery? it was presumed that, by being exposed to outside, infection and highly contaminated abscesses occurred.Was there any deficiency or anomaly of the interceed? if yes, please describe it.No further information is available.Was there any issue with interceed packaging? if yes, please describe it.No further information is available.Were cultures performed? results? no further information is available.Please provide details of wound injury? when was it occurred? no further information is available.What is physician¿s opinion as to the etiology of or contributing factors to this infection and abscess? it was presumed that, by being exposed to outside, infection and highly contaminated abscesses occurred.What is the current patient status? the patient is hospitalized.Sorry for late reply.No further information will be provided.
 
Event Description
It was reported that the patient underwent an ileus release operation/clean-contaminated operation for adhesive bowel obstruction on (b)(6) 2019 and just before abdominal closure, the absorbable adhesive barrier was placed on the pelvic floor, between the pelvic floor and the partially resected small intestine, where adhesions are expected and under the lower abdomen midline wound in front of the greater omentum.The bowel obstruction improved after the operation, but ct scan was performed on the 5th day because inflammation was prolonged.Also, wound infection occurred on the 3rd day, and heavy exudate was confirmed from the wound site.It was confirmed fist-sized abscesses on the pelvic floor and under the wound site by ct scan.By the result of the ct scan, emergency reoperation was performed on (b)(6) 2019.From the view of the reoperation, direct cause of the abscesses related to bowel obstruction surgery such as suture failure was not confirmed.The abscess under the abdominal wall is localized on the dorsal side of the wound on the ventral side of the omentum, and the abscess is a cream-colored pus with a bad odor, and the dorsal side of the abscess wall was made up of the gelled absorbable adhesive barrier.Fluid pooled between the device which was stuck to the abdominal wall, the back of the wound, and the peritoneum, and it leaked into the crevice between the wound.It was presumed that, by being exposed to outside, infection and highly contaminated abscesses occurred.The abscess on the pelvic floor was covered with the gelled absorbable adhesive barrier, and it was cloudy dark red colored liquid.The small intestine which had fallen into the pelvis due to the gelled absorbable adhesive barrier had been pushed out to the head side.This tumor is also presumed to be fluid abscess formed between the absorbable adhesive barrier stuck to the pelvic floor and the peritoneum.Both abscesses were cleaned, and drained, and gelled device was removed as much as possible.Originally, it was a semi-clean operation, but the wound was contaminated by re-operation, and it is now forced to manage as a highly contaminated wound.Although there was pus just below the wound, the contents of the pelvic floor were dark red liquid by absorbable adhesive barrier.The patient is still hospitalized with wound injury as it is open wound.The doctor opined that it was presumed that, by being exposed to outside, infection and highly contaminated abscesses occurred.No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
BARRIER, ABSORBABLE, ADHESION
Manufacturer (Section D)
ETHICON INC.
p.o. box 151, route 22 west
somerville NJ 08876 0151
Manufacturer (Section G)
ETHICON INC.-SAN LORENZO PR
982 road 183 km 8.3
san lorenzo PR 00754
Manufacturer Contact
darlene kyle
p.o. box 151, route 22 west
somerville, NJ 08876-0151
9082182792
MDR Report Key8891312
MDR Text Key154267132
Report Number2210968-2019-85517
Device Sequence Number1
Product Code MCN
UDI-Device Identifier10705031147843
UDI-Public10705031147843
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
P880047
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 07/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue Number4350XL
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 07/17/2019
Initial Date FDA Received08/13/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age41 YR
-
-