• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MELSUNGEN AG PERIFIX®; CATHETER, CONDUCTION, ANESTHETIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MELSUNGEN AG PERIFIX®; CATHETER, CONDUCTION, ANESTHETIC Back to Search Results
Catalog Number 4513258C
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as event seven of b.Braun melsungen ag internal report # (b)(4).No sample has been returned for investigation.Without the actual sample, a thorough investigation could not be performed and no specific conclusion can be drawn as to the cause of the reported event.If the sample and or additional pertinent information becomes available, a follow up report will be submitted.The product design is being updated to increase the force required to open the catheter connector under change control: hc-chc-m-div-1306.Note: this report is being filed for an item number that is not sold in the united states; however, this item or similar items are sold in the united sates by b.Braun medical, inc.
 
Event Description
As reported by the user facility (translation of user facility information by bbm sales organization in (b)(6)): event 7.Disconnection of the connector.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PERIFIX®
Type of Device
CATHETER, CONDUCTION, ANESTHETIC
Manufacturer (Section D)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, 34212
GM  34212
Manufacturer (Section G)
B. BRAUN MELSUNGEN AG
carl-braun-str. 1
melsungen, 34212
GM   34212
Manufacturer Contact
jonathan severino
861 marcon blvd.
allentown, PA 18109
4842408332
MDR Report Key8892280
MDR Text Key197858354
Report Number9610825-2019-00363
Device Sequence Number1
Product Code BSO
Combination Product (y/n)N
Reporter Country CodeCI
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number4513258C
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-