MEDTRONIC CRYOCATH LP ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER; PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION Back to Search Results
Model Number AFAPRO28
Device Problem Gas/Air Leak (2946)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/12/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that prior to a cryoablation procedure, it was not possible to get the balloon air empty during the flush of the balloon.The balloon catheter was replaced.There was no patient involvement.
 
Manufacturer Narrative
Product event summary: the balloon catheter, afapro28 with lot number 52933, was returned and analyzed.Visual inspection of the balloon catheter showed the catheter was intact with no apparent issues.Smart chip verification indicated the catheter was never used.Catheter passed the performance test and electrical integrity as per specification; impedance was also within specification.Insertion test was done through laboratory sheath and no issue was reproduced.Several flushes and aspirations were performed without issues while the catheter was inserted.In conclusion, the reported air ingress was not confirmed through testing.The balloon catheter passed the returned product inspection as per specification.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARCTIC FRONT ADVANCE PRO¿ CARDIAC CRYOABLATION CATHETER
Type of Device
PERCUTANEOUS CATHETER INTENDED FOR TREATMENT OF ATRIAL FIBRILLATION
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
MDR Report Key8893898
MDR Text Key154319509
Report Number3002648230-2019-00585
Device Sequence Number1
Product Code OAE
Combination Product (y/n)N
PMA/PMN Number
P100010
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date06/02/2021
Device Model NumberAFAPRO28
Device Catalogue NumberAFAPRO28
Device Lot Number52933
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/21/2019
Date Manufacturer Received08/30/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-