(b)(4).Investigation summary: examination of the returned device found the actuation feature to be jammed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
|
It was reported that the surgeon uses a combination of the ap measured sizer and the balancing sizer in his total knee arthroplasty's.He first sizes the knee with the measured sizer to get an idea of what size he's looking at, and then uses the balanced sizer to confirm.After dr.Finished using the balanced sizer completely and took it off, he noticed that the 2 pieces that make up the attune balanced sizer distractor rod seemed to be cross-threaded and could not be loosened back up.This did not delay the case in any way as surgeon was already finishing using it and continued to proceed to the next step.Replacements are needed to complete the femoral sizing and finishing tray.Upon examination, the internal rod component of the device was found to be jammed/frozen.
|