• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE BAL SIZER DISTRACTOR; ATTUNE INSTRUMENTS : SIZING/MEASURING INSTRUMENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE BAL SIZER DISTRACTOR; ATTUNE INSTRUMENTS : SIZING/MEASURING INSTRUMENTS Back to Search Results
Catalog Number 254400519
Device Problem Physical Resistance/Sticking (4012)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/19/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Investigation summary: examination of the returned device found the actuation feature to be jammed.Depuy synthes considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the surgeon uses a combination of the ap measured sizer and the balancing sizer in his total knee arthroplasty's.He first sizes the knee with the measured sizer to get an idea of what size he's looking at, and then uses the balanced sizer to confirm.After dr.Finished using the balanced sizer completely and took it off, he noticed that the 2 pieces that make up the attune balanced sizer distractor rod seemed to be cross-threaded and could not be loosened back up.This did not delay the case in any way as surgeon was already finishing using it and continued to proceed to the next step.Replacements are needed to complete the femoral sizing and finishing tray.Upon examination, the internal rod component of the device was found to be jammed/frozen.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE BAL SIZER DISTRACTOR
Type of Device
ATTUNE INSTRUMENTS : SIZING/MEASURING INSTRUMENTS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
Manufacturer (Section G)
DEPUY ORTHOPAEDICS, INC. 1818910
700 orthopaedic dr.
warsaw IN 46581 0988
Manufacturer Contact
kara ditty-bovard
700 orthopaedic dr.
warsaw, IN 46581-0988
6107428552
MDR Report Key8894432
MDR Text Key155691897
Report Number1818910-2019-101116
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295384403
UDI-Public10603295384403
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number254400519
Device Lot NumberABB25484
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/28/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/01/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Age71 YR
-
-