MPRI ATTAIN PERFORMA S MRI SURESCAN; DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE Back to Search Results
Model Number 459888
Device Problem Electrical /Electronic Property Problem (1198)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/11/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that a left ventricular (lv) lead polarity switch occurred a month ago.It was noted that some reprogramming was done, however another polarity switch occurred in the current month.It was also noted that there was invalid data for the event logs on the cardiac resynchronization therapy pacemaker (crt-p).The lv lead and the crt-p remain in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the lead was not returned for analysis, however, performance data collected from the device was received and analyzed.Analysis of the device memory indicated the impedance of the left ventricular pacing lead was beyond the expected upper range.Analysis of the device memory indicated the impedance trend of the left ventricular pacing lead was variable.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN PERFORMA S MRI SURESCAN
Type of Device
DRUG ELUTING PERMANENT LEFT VENTRICULAR (LV) PACEMAKER ELECTRODE
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8894652
MDR Text Key154322311
Report Number2649622-2019-14339
Device Sequence Number1
Product Code OJX
UDI-Device Identifier00643169601871
UDI-Public00643169601871
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P080006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 08/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/14/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date09/26/2020
Device Model Number459888
Device Catalogue Number459888
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/14/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/17/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
W4TR03 CRTP, 5076-52 LEAD, 5076-58 LEAD
Patient Outcome(s) Required Intervention;
Patient Age64 YR
-
-