Related manufacturer report number: 2135147-2019-00256.On (b)(6) 2019, an amplatzer muscular vsd occluder was selected for implant.The device was deployed but didn't fit the ventricular septal defect.The device was attempted to be retracted, but the occluder didn't collapse and retract into the delivery sheath.The tip of the delivery sheath kept folding in on itself.Meanwhile, the patient went into complete heart block and became hypotensive.The occluder and delivery system were withdrawn from the patient without collapsing the occluder.The procedure was cancelled and the patient was sent to the pediatric intensive care unit for continued monitoring.The heart block resolved without treatment.
|