• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BD KIESTRA LAB AUTOMATION BD KIESTRA¿ INOQULA+¿ TLA; MICROBIAL SPECIMEN INOCULATION/STREAKING INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BD KIESTRA LAB AUTOMATION BD KIESTRA¿ INOQULA+¿ TLA; MICROBIAL SPECIMEN INOCULATION/STREAKING INSTRUMENT Back to Search Results
Catalog Number 447213
Device Problems Nonstandard Device (1420); Use of Device Problem (1670); Electrical Power Problem (2925); Patient Data Problem (3197)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/18/2019
Event Type  malfunction  
Manufacturer Narrative
An anomaly was discovered in inoqula software (b)(4) that had the potential to cause a mismatch between a specimen and plate.The mismatch could happen when the following conditions occur together: inoqula hardware is not operating, such as during a power outage.Pipette has already drawn a specimen waiting to be dispensed on a plate, and user selects the ¿reset¿ function from the system menu.An immediate update to the software was made and validated.All impacted customers were provided with the update and information on any mis-associated data in their system.This was done via a recall and an 806 form was submitted on 5 august, 2019.Field correction initiated on 26july, 2019.Capa# (b)(4) has been opened to further investigate this issue.
 
Event Description
It was reported that while using a bd kiestra¿ inoqula+¿ tla to process patient urine samples, the customer reported discrepancies in bacterial growth between the blood agar plates and the macconkey agar plates being used for culture after a power outage.Following the discovery of the discrepancy, instrument log files uncovered a number of pipetting errors.During the investigation into this issue, through automated log-file analysis numerous mis-associated plates/specimens were discovered.Customers are being informed about the events, including ids and accession ids that were mis-associated.No erroneous results reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD KIESTRA¿ INOQULA+¿ TLA
Type of Device
MICROBIAL SPECIMEN INOCULATION/STREAKING INSTRUMENT
Manufacturer (Section D)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten
Manufacturer (Section G)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key8897865
MDR Text Key179303658
Report Number3010141591-2019-00118
Device Sequence Number1
Product Code JTC
UDI-Device Identifier00382904472139
UDI-Public00382904472139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 08/06/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/15/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number447213
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received07/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/11/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1119779-07/23/2019-004-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-