ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL
|
Back to Search Results |
|
Model Number 19AGFN-756 |
Device Problem
Fracture (1260)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 06/24/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
|
|
Event Description
|
On (b)(6) 2019, a double valve replacement was performed with a 19 mm regent heart valve in the aortic position.When the device was parachuted into the native annulus, the pivot guard area of the leaflet was fractured.All pieces were recovered.The device was exchanged for a 20 mm century medical inc ats bileaflet valve.No adverse consequences to the patient were reported, but the surgeon reported a clinically significant delay to the surgery.
|
|
Manufacturer Narrative
|
08/15/2019 mjr, supplemental report needed to address analysis.The reported event of one of the leaflets fractured and dislodged was confirmed.One leaflet had dislodged and fractured into two pieces.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the fractured leaflet.The cause of the reported event could not be conclusively determined, however it is consistent with damage caused by some external force applied to the leaflet and orifice, which overstressed the carbon material.
|
|