• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL PUERTO RICO, INC. SJM REGENT HEART VALVE W/FLEX CUFF; HEART-VALVE, MECHANICAL Back to Search Results
Model Number 19AGFN-756
Device Problem Fracture (1260)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 06/24/2019
Event Type  malfunction  
Manufacturer Narrative
Further information regarding this event has been requested.The results/method and conclusion codes along with investigation results will be provided in a subsequent submission.
 
Event Description
On (b)(6) 2019, a double valve replacement was performed with a 19 mm regent heart valve in the aortic position.When the device was parachuted into the native annulus, the pivot guard area of the leaflet was fractured.All pieces were recovered.The device was exchanged for a 20 mm century medical inc ats bileaflet valve.No adverse consequences to the patient were reported, but the surgeon reported a clinically significant delay to the surgery.
 
Manufacturer Narrative
08/15/2019 mjr, supplemental report needed to address analysis.The reported event of one of the leaflets fractured and dislodged was confirmed.One leaflet had dislodged and fractured into two pieces.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.There was no evidence of material defect in the carbon coating that may have caused or contributed to the fractured leaflet.The cause of the reported event could not be conclusively determined, however it is consistent with damage caused by some external force applied to the leaflet and orifice, which overstressed the carbon material.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SJM REGENT HEART VALVE W/FLEX CUFF
Type of Device
HEART-VALVE, MECHANICAL
Manufacturer (Section D)
ST. JUDE MEDICAL PUERTO RICO, INC.
p.o. box 998
lot 20 b st.
caguas, puerto rico 00725
MDR Report Key8898561
MDR Text Key154468043
Report Number2648612-2019-00060
Device Sequence Number1
Product Code LWQ
UDI-Device Identifier05414734005852
UDI-Public05414734005852
Combination Product (y/n)N
PMA/PMN Number
P810002
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 09/09/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date12/10/2022
Device Model Number19AGFN-756
Device Catalogue Number19AGFN-756
Device Lot Number6242786
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/08/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/18/2019
Initial Date FDA Received08/15/2019
Supplement Dates Manufacturer Received08/15/2019
Supplement Dates FDA Received09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
-
-