• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ST. JUDE MEDICAL, COSTA RICA LTDA DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, COSTA RICA LTDA DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER; CATHETER, INTRAVASCULAR, DIAGNOSTIC Back to Search Results
Model Number C408646
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 06/12/2019
Event Type  Injury  
Manufacturer Narrative
The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
During the procedure, a significant delay occurred.An error message appeared after the catheter was connected to the system.The device was reattempted three times, but the issue persisted and the procedure was delayed.Another device was used to complete the procedure with no patient consequences.
 
Manufacturer Narrative
One dragonfly optis imaging catheter was received for evaluation.A hoop was also returned.The results of the investigation concluded that an optical fiber fracture was detected at the splice region of the catheter shaft, consistent with the reported connection issue.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed.Based on the condition of the returned device and the information received, the cause of the optical fiber fracture remains unknown.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DRAGONFLY¿ OPTIS¿ KIT BOX IMAGING CATHETER
Type of Device
CATHETER, INTRAVASCULAR, DIAGNOSTIC
Manufacturer (Section D)
ST. JUDE MEDICAL, COSTA RICA LTDA
parque industrial, zona franca coyol s.a.
edificio #44b, calle 0, avenida 2, coyol
alajuela, costa rica 1897- 4050
CS  1897-4050
MDR Report Key8898844
MDR Text Key154527711
Report Number3008452825-2019-00415
Device Sequence Number1
Product Code DQO
UDI-Device Identifier00183739000654
UDI-Public00183739000654
Combination Product (y/n)N
PMA/PMN Number
K141769
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date03/31/2021
Device Model NumberC408646
Device Lot Number6971141
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/23/2019
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 07/26/2019
Initial Date FDA Received08/15/2019
Supplement Dates Manufacturer Received08/29/2019
Supplement Dates FDA Received09/10/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age80 YR
-
-