• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MEDICAL CORPORATION M SERIES; DEFIBRILLATOR/PACEMAKER Back to Search Results
Model Number MSERIES CCT
Device Problem Pacing Problem (1439)
Patient Problem No Information (3190)
Event Type  malfunction  
Manufacturer Narrative
Zoll medical corporation has received the product and will be providing a supplemental report when our investigation is completed.
 
Event Description
Complainant alleged that while attempting to pace a patient (age gender unknown) ,the device was unable to function as intended in the pacing mode.This is all the information available.Complainant did not indicate that there was any adverse effect to the patient due to the reported malfunction.
 
Manufacturer Narrative
The device was evaluated by zoll medical corporation.The reported malfunction was not duplicated or confirmed.The device was subjected to extensive troubleshooting which included full functionality testing without duplicating the reported malfunction.The device was recertified and returned to the customer.Its important to mention the device reacted to drop in hr which is normal.Its unknown why the drop in hr occurred and if it was symptomatic of the patience's condition.Analysis for reports of this type has not identified an increase in trend.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
M SERIES
Type of Device
DEFIBRILLATOR/PACEMAKER
Manufacturer (Section D)
ZOLL MEDICAL CORPORATION
269 mill road
chelmsford MA 01824
Manufacturer Contact
269 mill road
chelmsford, MA 01824
9784219552
MDR Report Key8903513
MDR Text Key154713999
Report Number1220908-2019-02327
Device Sequence Number1
Product Code DRO
Combination Product (y/n)N
PMA/PMN Number
K011865
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/16/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model NumberMSERIES CCT
Device Catalogue NumberM SERIES
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/30/2019
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received07/26/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured03/01/2005
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-