• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC NEUROSURGERY UNKNOWN STRATA VALVE/SHUNT; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC NEUROSURGERY UNKNOWN STRATA VALVE/SHUNT; SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS Back to Search Results
Model Number UNKNOWN-S
Device Problem Appropriate Term/Code Not Available (3191)
Patient Problems Edema (1820); Therapeutic Effects, Unexpected (2099)
Event Date 08/01/2019
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported the patient was implanted with a pressure regulator valve due to hydrocephalus.In (b)(6) 2019, there was a situation of edema that happened on the patient.The patient was hospitalized and underwent ct examination.The ct examination's results were showed to the surgeon, and the doctor said that the drainage was excessive.It was recommended to increase the patient's pressure by one level.The patient's pressure at discharge was 1.0.It was stated that if there was no change in the current pressure position, adjust to 1.5.Currently, the patient was hospitalized.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Additional information received reported that on august 19, 2019, the manufacturer representative adjusted the pressure for the patient at the hospital, and it resolved the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN STRATA VALVE/SHUNT
Type of Device
SHUNT, CENTRAL NERVOUS SYSTEM AND COMPONENTS
Manufacturer (Section D)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer (Section G)
MEDTRONIC NEUROSURGERY
125 cremona drive
goleta CA 93117
Manufacturer Contact
stacy ruemping
7000 central avenue ne rcw215
minneapolis, MN 55432
7635260594
MDR Report Key8907869
MDR Text Key154839513
Report Number2021898-2019-00313
Device Sequence Number1
Product Code JXG
Combination Product (y/n)N
Reporter Country CodeCH
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 08/26/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberUNKNOWN-S
Device Catalogue NumberUNKNOWN-S
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/16/2019
Initial Date FDA Received08/19/2019
Supplement Dates Manufacturer Received08/21/2019
Supplement Dates FDA Received08/26/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization;
Patient Age71 YR
Patient Weight60
-
-