Catalog Number LG-03000 |
Device Problem
Break (1069)
|
Patient Problem
No Patient Involvement (2645)
|
Event Date 07/31/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).The device has not been returned for investigation.Teleflex will continue to monitor and trend related events.
|
|
Event Description
|
It was reported that of the last four kits opened, two of them the medicine vials were broken.
|
|
Manufacturer Narrative
|
(b)(4).A device history record review was performed with a potentially relevant finding.A nonconformance was initiated only as a general nc to temporarily allow certain finished good part numbers to be packaged with additional gauze in order to better secure medication ampules in the inner tray.Complaint verification testing could not be performed as no sample was returned for analysis.A device history record review was performed with a potentially relevant finding.Based on the information provided, the potential root cause of this issue is design related.A capa was initiated to further investigate this complaint issue.
|
|
Event Description
|
It was reported that of the last four kits opened, two of them the medicine vials were broken.
|
|