Model Number IST-501
Device Problem False Negative Result (1225)
Patient Problem No Information (3190)
Event Date 07/11/2019
Event Type  Injury  
Manufacturer Narrative
Results pending completion of investigation.
 
Event Description
It was reported that on (b)(6) 2019, a patient presented to the facility with symptoms of a sore throat, fever, coughing and a headache.The patient had a strep a kit administered without a timer and a negative result occurred.On (b)(6) 2019 a confirmatory test was performed however no additional information was provided.A negative patient outcome was reported with no other information received although requested.Due to the unknown severity of the negative patient outcome being reported in this case, a conservative approach wil be taken and a serious injury mdr is filed.Per device labeling "limitations" section, "a negative result obtained from this kit should be confirmed by culture.A negative result may be obtained if the concentration of the strep a antigen present in the throat swab is not adequate or is below the detectable level of the test.As with all diagnostic tests, all results must be interpreted together with other clinical information available to the physician." troubleshooting occurred with a discussion about possible causes for unexpected results focusing on proper sampling technique and storage conditions per package insert.
 
Manufacturer Narrative
This was only one event although uncertain which lot number was associated with the event.The initial filing had lot number sta8120004, but the other potentially involved lot numbers are sta8090042, sta8050042, sta8110032, sta8120010, sta9010014, sta9010055, sta8090043, sta8110027.
 
Manufacturer Narrative
Investigation conclusion representing lot numbers sta8120004, sta8090042, sta8050042, sta8110032, sta8120010, sta9010014, sta9010055, sta8090043, sta8110027: an investigation was performed on retention and returned products for these reported lot numbers.Retention and returned devices were tested with qc cut-off positive sample (2.5e07org/ml) and results were read at 5 minutes.All devices produced expected positive results at the read time.No false negative results were obtained during in-house testing.The case details were reviewed along with the complaint history for the reported issue and no indications of a systemic issue were identified.Batch record review found no relevant non-conformances; the lot met all final release specifications.It was reported two drops of reagent a and two drops of reagent b were added into the extraction tube and no color change was observed.The swab was swirled 5-10 times or for 15 seconds and then left in the solution for two minutes.A timer was not used.Per the package insert: add four full drops of reagent 1 (red in color) and four full drops of reagent 2 (colorless) into the extraction tube.The addition of reagent 2 to reagent 1 changes the color of the solution from red to pale yellow.Tap the bottom of the tube gently to mix the liquid.Immediately add the throat swab into the tube of pale yellow solution.Rotate the swab vigorously 10 times in the tube.Leave the swab in the tube for 1 minute.Deviations in testing technique cannot be ruled out as a root cause for the reported issue.A negative result obtained from this kit should be confirmed by culture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSULT STREP A DIPSTICK 5003 25T
Type of Device
STREP A TEST
Manufacturer (Section D)
ALERE SAN DIEGO
9995 summers ridge rd
san diego CA 92121
MDR Report Key8918993
MDR Text Key155194528
Report Number2027969-2019-00240
Device Sequence Number1
Product Code GTY
UDI-Device Identifier20612479202737
UDI-Public(01)20612479202737(17)201130(10)STA8120004
Combination Product (y/n)N
PMA/PMN Number
K010582
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup,Followup
Report Date 10/30/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Device Operator Health Professional
Device Expiration Date11/30/2020
Device Model NumberIST-501
Device Lot NumberMULTIPLE LOT NUMBERS-SEE H10
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/24/2019
Initial Date Manufacturer Received 07/31/2019
Initial Date FDA Received08/21/2019
Supplement Dates Manufacturer Received07/30/2019
10/15/2019
Supplement Dates FDA Received08/22/2019
10/31/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age17 YR
Patient Weight80
-
-