Catalog Number UNK HIP FEMORAL HEAD
Device Problems Loss of Osseointegration (2408); Naturally Worn (2988)
Patient Problems Inflammation (1932); Pain (1994); Tissue Damage (2104); Inadequate Osseointegration (2646); No Code Available (3191)
Event Date 03/14/2017
Event Type  Injury  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Painful post-op.This case is from health authority.It was reported that the patient was with right hip osteoarthritis, so had undergone the operation of right hip replacement on (b)(6) 2019 with asr products.In 2016, the patient felt painful, the pain was gradually aggravated on the operated joint, and crutches were needed for the movement.The patient was diagnosed with aseptic loosening of prosthesis, and was undergone right hip revision surgery on (b)(6) 2017.During the surgery, the surgeon noted metal-on-metal wear, metal ion result in soft tissue inflammation, and prosthesis loosening.The patient's joint function recovered after revision surgery.
 
Manufacturer Narrative
Product complaint # (b)(4).Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provision of 21 cfr, part 803.The report may be based on the information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review was not possible because the required lot code(s) was not provided.Based on previous investigations this complication of joint replacement is unlikely to have been the result of a device failing to meet required specifications.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Corrective action was not indicated.Device history lot : null.Device history batch : null.Device history review : null.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN HIP FEMORAL HEAD
Type of Device
HIP FEMORAL HEAD
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8 DT
UK  LS11 8 DT
MDR Report Key8922327
MDR Text Key155261483
Report Number1818910-2019-102022
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,o
Remedial Action Recall
Type of Report Initial,Followup,Followup
Report Date 08/01/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/01/2019
Initial Date FDA Received08/22/2019
Supplement Dates Manufacturer Received08/01/2019
10/22/2019
Supplement Dates FDA Received08/28/2019
10/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age63 YR
-
-