• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SIEMENS AG / SIEMENS HEALTHCARE GMBH SIEMENS MRI; SYSTEM, NUCLEAR, MAGNETIC RESONANCE IMAGING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SIEMENS AG / SIEMENS HEALTHCARE GMBH SIEMENS MRI; SYSTEM, NUCLEAR, MAGNETIC RESONANCE IMAGING Back to Search Results
Model Number 7391167
Device Problem Electromagnetic Interference (1194)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/01/2019
Event Type  Injury  
Event Description
Ferrous non-mri safe rolling iv pole was brought into mri room and was immediately pulled into magnet.There was no injury to pt or staff.Magnet was powered down via controlled process by field serviceman retaining helium level.Unit was returned to full service august 2, 2019.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIEMENS MRI
Type of Device
SYSTEM, NUCLEAR, MAGNETIC RESONANCE IMAGING
Manufacturer (Section D)
SIEMENS AG / SIEMENS HEALTHCARE GMBH
MDR Report Key8922677
MDR Text Key155341612
Report NumberMW5089239
Device Sequence Number1
Product Code LNH
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Voluntary
Reporter Occupation Physician
Type of Report Initial
Report Date 08/02/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number7391167
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received Not provided
Initial Date FDA Received08/21/2019
Was Device Evaluated by Manufacturer? No Information
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-