• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD MODIFIED SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A1059
Device Problems Mechanical Problem (1384); Device Slipped (1584)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Linked to mfg report number: 3004608878-2019-00219, 3004608878-2019-00221.
 
Event Description
This is 1 of 3 reports.A customer reported that the a1059 mayfield modified skull clamp had a "richet" tightening/slipping down problem.The date of the event was not specified.According to the customer, they did a preventive maintenance check and there was no patient involvement.Additional information has been requested.
 
Event Description
N/a.
 
Manufacturer Narrative
Device identifier # 10381780253457.The unit was received with the ratchet extension arm not going through the base smoothly.The lock has rotational and lateral movement.The unit is less than a year old.The device history record review showed no abnormalities related to the reported failure.The device passed all required inspection points with no associated mrr¿s, variances or rework.The reported complaint was confirmed from the evaluation of item.The returned unit did not meet all specific functional test.The complaint was likely caused by wear and tear.The definite root cause could not be reliably determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD MODIFIED SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key8925170
MDR Text Key187117737
Report Number3004608878-2019-00220
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
PRE-AMEND
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Type of Report Initial,Followup
Report Date 08/05/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/23/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA1059
Device Lot Number194
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Date Manufacturer Received09/09/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-