Catalog Number A1059 |
Device Problems
Mechanical Problem (1384); Device Slipped (1584)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
The device was not yet returned to the manufacturer for analysis.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.Linked to mfg report number: 3004608878-2019-00219, 3004608878-2019-00221.
|
|
Event Description
|
This is 1 of 3 reports.A customer reported that the a1059 mayfield modified skull clamp had a "richet" tightening/slipping down problem.The date of the event was not specified.According to the customer, they did a preventive maintenance check and there was no patient involvement.Additional information has been requested.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Device identifier # 10381780253457.The unit was received with the ratchet extension arm not going through the base smoothly.The lock has rotational and lateral movement.The unit is less than a year old.The device history record review showed no abnormalities related to the reported failure.The device passed all required inspection points with no associated mrr¿s, variances or rework.The reported complaint was confirmed from the evaluation of item.The returned unit did not meet all specific functional test.The complaint was likely caused by wear and tear.The definite root cause could not be reliably determined.
|
|