• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. TEMPERATURE SENSING FOLEY CATHETER; FOLEY CATHETER (LATEX)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX® I.C. TEMPERATURE SENSING FOLEY CATHETER; FOLEY CATHETER (LATEX) Back to Search Results
Catalog Number 129414M
Device Problems Deflation Problem (1149); Inflation Problem (1310); Improper Flow or Infusion (2954)
Patient Problems No Patient Involvement (2645); No Known Impact Or Consequence To Patient (2692)
Event Date 07/20/2019
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the balloon did not inflate due to the water that flowed out from the tip of the catheter during a pretest.
 
Event Description
It was reported that the balloon did not inflate due to the water that flowed out from the tip of the catheter during a pretest.
 
Manufacturer Narrative
The reported event was unconfirmed as the problem could not be reproduced.Evaluation of the device found no leakage observed even when pressure was applied on the sac.The sample underwent a 7 day leak test and no leakage was observed.There was no leakage from the valve.The catheter was dissected and no conditions were found that could have contributed to the reported event.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "[warnings] 1.Method for use (1)do not inflate the balloon in the urethra.[the urethra may be injured.] (2)do not pull the catheter hard.[the bladder/urethra may be injured.] 2.Applicable patients ·patients with delirium who might pull out catheter [when patient tugs at catheter unconsciously, the bladder and urethra may be damaged.] [contraindications] 1.Method for use: (1) do not reuse.(2) do not resterilize.(3) this device contains 10% povidone-iodine.For patients with past history of allergic hypersensitivity to povidone-iodine or iodine, consider using alternative disinfectants.(4) be careful that the catheter is notexposed to ointments, contrast medium or oil-based lubricants (including vegetable oils such as olive oil, mineral oils such as white petrolatum and animal oils).[they may damage the device and may burst balloon.] (5) do not hold the device with forceps, etc.Avoid contact with any blades or sharp-edged instruments.[catheter damage may cause balloon rupture and accidental balloon removal or failure to deflate or remove the balloon.] 2.Applicable patients patients with known allergy to silver coated catheter" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX® I.C. TEMPERATURE SENSING FOLEY CATHETER
Type of Device
FOLEY CATHETER (LATEX)
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington GA 30014
MDR Report Key8925220
MDR Text Key193983010
Report Number1018233-2019-05056
Device Sequence Number1
Product Code MJC
Combination Product (y/n)N
PMA/PMN Number
K910318
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Type of Report Initial,Followup
Report Date 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date10/28/2022
Device Catalogue Number129414M
Device Lot Number9CA154
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/09/2019
Initial Date Manufacturer Received 07/31/2019
Initial Date FDA Received08/23/2019
Supplement Dates Manufacturer Received08/28/2019
Supplement Dates FDA Received09/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-