Catalog Number 447213
Device Problems Nonstandard Device (1420); Data Problem (3196)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 02/22/2019
Event Type  malfunction  
Manufacturer Narrative
Field correction initiated on 26july, 2019.Capa# (b)(4) has been opened to further investigate this issue.An anomaly was discovered in inoqula software (b)(4) that had the potential to cause a mismatch between a specimen and plate.The mismatch could happen when the following conditions occur together: inoqula hardware is not operating, such as during a power outage.Pipette has already drawn a specimen waiting to be dispensed on a plate, and user selects the ¿reset¿ function from the system menu.An immediate update to the software was made and validated.All impacted customers were provided with the update and information on any mis-associated data in their system.This was done via a recall and an 806 form was submitted on 5 august, 2019.
 
Event Description
It was reported that while using a bd kiestra¿ inoqula+¿ tla an anomaly is present in inoqula/ inoqula+ software version (b)(4).During the investigation into this issue, through automated log-file analysis mis-associated plates/specimens were discovered.Customers have been informed about the events, including ids and accession ids that were mis-associated.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD KIESTRA¿ INOQULA+¿ TLA
Type of Device
MICROBIAL SPECIMEN INOCULATION/STREAKING INSTRUMENT
Manufacturer (Section D)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten
Manufacturer (Section G)
BD KIESTRA LAB AUTOMATION
6 marconilaan
drachten
Manufacturer Contact
brett wilko
9450 south state street
sandy, UT 84070
8015652341
MDR Report Key8932534
MDR Text Key181648057
Report Number3010141591-2019-00174
Device Sequence Number1
Product Code JTC
UDI-Device Identifier00382904472139
UDI-Public00382904472139
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Reporter Occupation Other Health Care Professional
Remedial Action Recall
Type of Report Initial
Report Date 08/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number447213
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 07/30/2019
Initial Date FDA Received08/26/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/05/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Removal/Correction Number1119779-07/23/2019-004-C
Patient Sequence Number1
Patient Outcome(s) Other;
-
-