• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SàRL CH SI POLYAXL SCREW 6 X 45MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SàRL CH SI POLYAXL SCREW 6 X 45MM; ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE Back to Search Results
Model Number 179712645
Device Problem Loose or Intermittent Connection (1371)
Patient Problem No Code Available (3191)
Event Type  Injury  
Manufacturer Narrative
(b)(4).The device is available for evaluation.Investigation will be conducted.Follow up will be filed with the investigation results.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that on an unknown date, after an unknown procedure, the patient implants migrated posteriorly after a fall.The patient went to the clinic on may 8 and notified the surgeon about the fall while taking out the trash cans.Unknown date on the fall.The surgeon assumed that the cage migrated and had a ct scan done to confirm.According to the surgeon, the implants that are in the patient are not concorde lift but was possibly concorde bullet proti.There was no revision confirmation and patient symptoms are unknown.This complaint involves one (1) device.
 
Manufacturer Narrative
Product complaint (b)(4).Udi (b)(4).Device was not returned for evaluation.A review of the device history record was conducted.No issues were identified during the manufacturing and release of this product that could have contributed to the problem reported by the customer.Without the return of the device, we are unable to confirm the reported issue or identify the root cause.As there has been no issue identified in the manufacturing or release of the device that could have contributed to the problem reported by the customer, this complaint file will be closed with no further action required.Should more information and/or the sample be provided at a later time, this complaint will be reopened and device evaluated.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SI POLYAXL SCREW 6 X 45MM
Type of Device
ORTHOSIS, SPINAL PEDICLE FIXATION, FOR DEGENERATIVE DISC DISEASE
Manufacturer (Section D)
MEDOS INTERNATIONAL SàRL CH
chemin-blanc 38
le locle 02400
SZ  02400
Manufacturer (Section G)
MEDOS INT SPINE
chemin blanc 38
le locle
SZ  
Manufacturer Contact
kara ditty-bovard
325 paramount drive
raynham, MA 02767
6103142063
MDR Report Key8934438
MDR Text Key155965829
Report Number1526439-2019-52038
Device Sequence Number1
Product Code NKB
UDI-Device Identifier10705034067735
UDI-Public(01)10705034067735
Combination Product (y/n)N
PMA/PMN Number
K160904
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 05/08/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number179712645
Device Catalogue Number179712645
Device Lot Number157585
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/12/2019
Initial Date FDA Received08/27/2019
Supplement Dates Manufacturer Received09/17/2019
Supplement Dates FDA Received09/18/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/29/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Weight125
-
-