Model Number D314TRG
Device Problems Premature Discharge of Battery (1057); Delayed Charge Time (2586)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/09/2019
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: 511211 oscor medical lead, implanted: (b)(6) 2005.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the cardiac resynchronization therapy defibrillator (crt-d) was known to have reached recommended replacement time (rrt), but now exhibited a warning alerts observing charge circuit time-out (cto), long charge circuit time of 30 seconds, and charge circuit inactive (cti).Premature battery depletion was suspect with less than normal elective replacement indicator (eri) longevity.The device was explanted and replaced.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Product event summary: the device was returned and analyzed.Analysis of the device found high internal battery resistance.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROTECTA XT CRT-D
Type of Device
DEFIBRILLATOR, AUTOMATIC IMPLANTABLE CARDIOVERTER, WITH CARDIAC RESYNCHRONIZATIO
Manufacturer (Section D)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC MED REL MEDTRONIC PUERTO RICO
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8935551
MDR Text Key155664114
Report Number3004209178-2019-16424
Device Sequence Number1
Product Code NIK
UDI-Device Identifier00613994677884
UDI-Public00613994677884
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010031
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 12/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/14/2014
Device Model NumberD314TRG
Device Catalogue NumberD314TRG
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/05/2019
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 08/09/2019
Initial Date FDA Received08/27/2019
Supplement Dates Manufacturer Received12/02/2019
Supplement Dates FDA Received12/04/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/16/2013
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Treatment
694965 LEAD, 5076-52 LEAD
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age63 YR
-
-