Model Number X SERIES |
Device Problem
Unable to Obtain Readings (1516)
|
Patient Problem
No Patient Involvement (2645)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Zoll medical corporation has not received the product for evaluation and this complaint is still under investigation.
|
|
Event Description
|
Complainant alleged that during biomed testing, the device displayed a "defib pad short" message.Complainant indicated that there was no patient involvement in the reported malfunction.
|
|
Manufacturer Narrative
|
A zoll representative went on site and evaluated the device.During evaluation the representative observed the device was being used to attempt a 30j shock test while the mfc is connected to an unopened cpr stat-padz.Cpr stat-padz does not contain a shorting wire, therefore it cannot be used to perform a manual 30j self-test by design.It will cause the device to not see a short and will prompt the reported pads/paddles lead fault message.This is not an indication of a device malfunction.Therefore this claim is being closed as device meets specification.The device was recertified.No trend is associated with reports of this type.
|
|