The reported issue was confirmed.The device was returned for evaluation.Visual inspection found the information printed on the cap of the catheter was not clear and smeared.No other components were returned together with this sample.Temper evidence seal also have been remove and not returned together with the sample.Unable to determine how and when the problem occurred.This occurred post manufacturing process.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: ¿[shape, configuration and principles] bard®i.C.Silver foley tray b consists of a balloon catheter, urine-collecting bag for closed drainage system, syringe prefilled with sterile water, water-soluble lubricant, antiseptic solution, tweezers, waterproof sheet, gauze pads, cotton balls and vinyl gloves.There are several types of closed drainage bags.The bag included in the tray will depend on the product. available in sizes 8 to 24 every 2fr" h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.
|