AGA MEDICAL CORPORATION AMPLATZER SEPTAL OCCLUDER; TRANSCATHETER SEPTAL OCCLUDER Back to Search Results
Model Number 9-ASD-038
Device Problem Migration or Expulsion of Device (1395)
Patient Problem Nausea (1970)
Event Date 08/11/2019
Event Type  Injury  
Manufacturer Narrative
An event of embolization was reported.The results of the investigation are inconclusive since the device was not returned for analysis.The device history record was reviewed to ensure that each manufacturing and inspection operation was performed and the product met all specifications at the time of commercialization.Based on the information received, the cause of the reported incident could not be conclusively determined.
 
Event Description
A device was placed on 8/6.Upon post echo 24 hours later the device was in good position.On 8/11 the patient reported nausea and sweating and, self reported to the emergency room with reported symptom relief shortly after.On echo in the ed, the device was noted to be in the la.An attempt was made to retrieve the device transcatheter unsuccessfully.After discussion with the team the decision was made to anticoagulate the patient and go to the or for device removal and surgical closure in the coming days.The patient remained stable throughout the procedure and was admitted to the hospital awaiting surgery.
 
Event Description
On (b)(6) 2019, a 38mm amplatzer septal occluder was implanted.On (b)(6) 2019, the patient presented to the emergency room with complaints of nausea and sweating.An echocardiogram was performed which revealed the device had embolized to the la.The device was unsuccessfully retrieved via transcatheter.On (b)(6) 2019, the patient was transferred to the operating room where the device was surgically removed and the asd surgically repaired.The patient is reported to be stable.The user alleges the device embolization was likely due to the large size of the atrial septal defect.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AMPLATZER SEPTAL OCCLUDER
Type of Device
TRANSCATHETER SEPTAL OCCLUDER
Manufacturer (Section D)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer (Section G)
AGA MEDICAL CORPORATION
5050 nathan lane north
plymouth MN 55442
Manufacturer Contact
stephanie o' sullivan
5050 nathan lane north
plymouth, MN 55442
6517565400
MDR Report Key8942903
MDR Text Key155977363
Report Number2135147-2019-00270
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00811806010267
UDI-Public00811806010267
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P000039
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial,Followup
Report Date 09/17/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/31/2019
Device Model Number9-ASD-038
Device Catalogue Number9-ASD-038
Device Lot Number4791270
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/11/2019
Initial Date FDA Received08/28/2019
Supplement Dates Manufacturer Received08/19/2019
Supplement Dates FDA Received09/17/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/17/2014
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age62 YR
Patient Weight92
-
-