• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BAXTER HEALTHCARE - ROUND LAKE EM 2400, LOAD CELL MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - ROUND LAKE EM 2400, LOAD CELL MODULE; SYSTEM/DEVICE, PHARMACY COMPOUNDING Back to Search Results
Catalog Number 2400L
Device Problem Break (1069)
Patient Problem No Patient Involvement (2645)
Event Date 08/07/2019
Event Type  malfunction  
Manufacturer Narrative
The device was received for evaluation.External and internal inspections were performed and revealed that the load cell cable assembly had torn insulation.Functional testing failed due to an inoperative load cell.The reported condition was verified.The cause of the condition was determined to be a damaged load cell cable assembly.To resolve the issue, the load cell cable assembly will be replaced.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that the load cell of an exactamix automated compounding device was not working.The customer stated that the end wire was crimped (further described as compressed, bent) where it connects to the load cell.The customer also stated that the wires were exposed.This issue was noted upon power up.There was no patient involvement.No additional information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EM 2400, LOAD CELL MODULE
Type of Device
SYSTEM/DEVICE, PHARMACY COMPOUNDING
Manufacturer (Section D)
BAXTER HEALTHCARE - ROUND LAKE
round lake IL
Manufacturer (Section G)
BAXTER HEALTHCARE - ROUND LAKE
25212 w. illinois route 120
round lake IL 60073
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8946910
MDR Text Key157871507
Report Number1416980-2019-04715
Device Sequence Number1
Product Code NEP
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 08/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received08/29/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number2400L
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/13/2019
Date Manufacturer Received08/07/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-