• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: B. BRAUN MEDICAL INC APEX¿; COMPOUNDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

B. BRAUN MEDICAL INC APEX¿; COMPOUNDER Back to Search Results
Model Number AX1000
Device Problem Application Program Problem: Dose Calculation Error (1189)
Patient Problem No Patient Involvement (2645)
Event Date 08/18/2019
Event Type  malfunction  
Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device logs and the compounding activity report (car) have been received.The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
 
Event Description
As per user facility: customer reported issues with incorrect order volumes on the compounding activity report from one of the orders compounded.The order volume, delivered volume, expected weight and total delivered volume of individual ingredients are incorrect.No patient involvement.
 
Manufacturer Narrative
This report has been identified as b.Braun medical inc.Internal report number (b)(4).The device logs and the compounding activity report (car) have been received and reviewed.This is a known software issue.The software issue has to do with scanning the source line barcode from the main screen to bring up the "replace container" screen for the current universal ingredient (ui).When this is done in the middle of an order (only possible if user first presses "pause"), then the universal ingredient will revert from whatever station it was (2 or 3) to station 1.The final flush of the order will be performed with sterile water from station 1 instead of the intended universal ingredient.This will result in less volume of the intended universal ingredient in the amount of the flush volume and more of station 1 (water) than intended in the amount of the flush volume.Some of this flush volume will not make it into the final container, as it will be used to fill the manifold for the next order.Also, a manifold flush will be required prior to that next order.Additionally, the compounding activity report will show incorrect values for both delivered volume and ordered volume, neither will reflect the intended ordered volume.There is an active apex software project to address this issue.If additional pertinent information becomes available a follow-up report will be filed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
APEX¿
Type of Device
COMPOUNDER
Manufacturer (Section D)
B. BRAUN MEDICAL INC
1601 wallace drive, suite 150
carrollton TX 75006 6690
MDR Report Key8955182
MDR Text Key158637373
Report Number1641965-2019-00013
Device Sequence Number1
Product Code NEP
UDI-Device Identifier04046955048502
UDI-Public(01)04046955048502
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 10/29/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberAX1000
Device Catalogue NumberAX1000
Was Device Available for Evaluation? No
Date Manufacturer Received08/19/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-