• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC CRYOCATH LP FLEXCATH ADVANCE STEERABLE SHEATH; CATHETER, STEERABLE Back to Search Results
Model Number 4FC12
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Perforation of Vessels (2135); Cardiac Tamponade (2226)
Event Date 08/09/2019
Event Type  Injury  
Manufacturer Narrative
Product event summary: the data files were returned and analyzed.The data files showed at least eight applications were performed with balloon catheter 2af284 with lot 11346 and system notice 50008 ¿the system has detected a software error and stopped the injection¿ was triggered in the first application.In conclusion, the reported cardiac tamponade could not be confirmed through testing.There is no indication of relation of adverse event to the performance of the cryo device.If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that during a cryo ablation procedure, the patients vitals signs decreased and cardiopulmonary resuscitation was initiated.A cardiac tamponade was noted and pericardiocentesis was performed.The patient required surgical intervention to repair a tear to the pulmonary vein resulting in an extended hospital stay and patient remains in the hospital on ventilation support with steady improvements.The case was aborted.No further patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FLEXCATH ADVANCE STEERABLE SHEATH
Type of Device
CATHETER, STEERABLE
Manufacturer (Section D)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA  H9R 5Z8
Manufacturer (Section G)
MEDTRONIC CRYOCATH LP
9000 autoroute transcanadienne
pointe-claire,qc H9R 5 Z8
CA   H9R 5Z8
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key8956234
MDR Text Key156408916
Report Number3002648230-2019-00635
Device Sequence Number1
Product Code DRA
UDI-Device Identifier00763000043773
UDI-Public00763000043773
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K123591
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/03/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date01/30/2021
Device Model Number4FC12
Device Catalogue Number4FC12
Device Lot Number63607
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/09/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/01/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Life Threatening; Required Intervention;
Patient Age73 YR
-
-