Product event summary: the data files were returned and analyzed.The data files showed at least eight applications were performed with balloon catheter 2af284 with lot 11346 and system notice 50008 ¿the system has detected a software error and stopped the injection¿ was triggered in the first application.In conclusion, the reported cardiac tamponade could not be confirmed through testing.There is no indication of relation of adverse event to the performance of the cryo device.If information is provided in the future, a supplemental report will be issued.
|