• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

W.L. GORE & ASSOCIATES GORE® CARDIOFORM SEPTAL OCCLUDER; TRANSCATHETER, SEPTAL OCCLUDER Back to Search Results
Catalog Number GSX0025A
Device Problem Positioning Problem (3009)
Patient Problem Embolism (1829)
Event Date 08/14/2019
Event Type  Injury  
Event Description
It was reported on 08/13/2019, the physician implanted a 25 mm gore® cardioform septal occluder to close a patent foramen ovale.No balloon stop flow sizing was used.The following day, at the discharge echocardiography, the device was noted to have embolized to the descending aorta.It was not obstructing blood flow, so the device was removed with a snare on (b)(6) 2019 with no issues.The defect was closed with a competitor device.The patient was doing well following the procedure.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GORE® CARDIOFORM SEPTAL OCCLUDER
Type of Device
TRANSCATHETER, SEPTAL OCCLUDER
Manufacturer (Section D)
W.L. GORE & ASSOCIATES
flagstaff AZ
Manufacturer (Section G)
KENDRICK PEAK MPD B/P
4250 w. kiltie lane
flagstaff AZ 86005
Manufacturer Contact
dan kitterman
1500 n. 4th street
9285263030
MDR Report Key8956630
MDR Text Key160914414
Report Number2017233-2019-00784
Device Sequence Number1
Product Code MLV
UDI-Device Identifier00733132631025
UDI-Public00733132631025
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P050006
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Physician
Type of Report Initial
Report Date 08/14/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/29/2021
Device Catalogue NumberGSX0025A
Device Lot Number20533081
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received Not provided
Initial Date FDA Received09/03/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/30/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age21 YR
Patient Weight48
-
-