• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER SPINE-US XIA 3 TITANIUM BLOCKER; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER SPINE-US XIA 3 TITANIUM BLOCKER; THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM Back to Search Results
Catalog Number 48230000
Device Problems Loose or Intermittent Connection (1371); Unstable (1667); Migration (4003)
Patient Problem Injury (2348)
Event Date 05/05/2019
Event Type  Injury  
Event Description
It was reported that a patient presented with four loose xia titanium blockers which, ¿generated instability in the spine.¿ this record represents the fourth of the four blockers.
 
Manufacturer Narrative
Visual, dimensional, material and functional analysis could not be performed as the device was not returned.Device history records and complaint history records were reviewed.No relevant manufacturing issues were identified as all units met stryker specifications and no similar complaints were identified.From the xia ifu: the surgeon must warn the patient that the devices cannot and do not replicate the flexibility, strength, reliability or durability of normal healthy bone, that the implants can break or become damaged as a result of strenuous activity or trauma, and that the devices may need to be replaced in the future.If the patient is involved in an occupation or activity which applies inordinate stress upon the implant (e.G., substantial walking, running, lifting, or muscle strain) the surgeon must advise the patient that resultant forces can cause failure of the devices.Patients who are overweight may be responsible for additional stresses and strains on the device which can speed up metal fatigue and/or lead to deformation or failure of the implants.The size and shape of the bone structures determine the size, shape and type of the implants.Once implanted, the implants are subjected to stresses and strains.These repeated stresses on the implants should be taken into consideration by the surgeon at the time of the choice of the implant, during implantation as well as in the post-operative follow-up period.Indeed, the stresses and strains on the implants may cause metal fatigue or fracture or deformation of the implants, before the bone graft has become completely consolidated.This may result in further side effects or necessitate the early removal of the osteosynthesis device.The devices currently cannot be located by the site, so evaluation on the implants could not be performed.Without evaluating the device, a definite root cause cannot be determined.Possible root causes include over/under tightening of the blocker, off loading of the blocker, and/or excessive post op activity.
 
Event Description
It was reported that a patient presented with four loose xia titanium blockers which, ¿generated instability in the spine.¿ this record represents the fourth of the four blockers.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XIA 3 TITANIUM BLOCKER
Type of Device
THORACOLUMBOSACRAL PEDICLE SCREW SYSTEM
Manufacturer (Section D)
STRYKER SPINE-US
2 pearl court
allendale NJ 07401
MDR Report Key8957979
MDR Text Key156403226
Report Number0009617544-2019-00107
Device Sequence Number1
Product Code NKB
UDI-Device Identifier04546540560193
UDI-Public04546540560193
Combination Product (y/n)N
PMA/PMN Number
K142381
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 11/21/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number48230000
Device Lot Number72C
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 05/15/2019
Initial Date FDA Received09/03/2019
Supplement Dates Manufacturer Received10/25/2019
Supplement Dates FDA Received11/21/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-