Additional information : the device was returned for evaluation.A visual inspection was performed, and all components were correctly placed and according to specifications (no visual defect were observed).The sample was then functionally tested and primed, and it was determined that the device primed and flowed normally.Additionally, the filter was pressure tested with no issue noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
|