MAUDE Adverse Event Report: BAXTER HEALTHCARE - DOMINICAN REPUBLIC MICRO-VOLUME EXTENSION SET; FILTER, INFUSION LINE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BAXTER HEALTHCARE - DOMINICAN REPUBLIC MICRO-VOLUME EXTENSION SET; FILTER, INFUSION LINE Back to Search Results
Catalog Number 2N3350
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Type  malfunction  
Manufacturer Narrative
This event occurred on an unspecified date in (b)(6) 2019.Should additional relevant information become available, a supplemental report will be submitted.
 
Event Description
It was reported that an unspecified quantity of non-dehp extension set leaked at the filter.The event occurred during an infusion with an unspecified medication.There was no patient injury or medical intervention associated with this event.No additional information is available.
 
Manufacturer Narrative
Additional information : the device was returned for evaluation.A visual inspection was performed, and all components were correctly placed and according to specifications (no visual defect were observed).The sample was then functionally tested and primed, and it was determined that the device primed and flowed normally.Additionally, the filter was pressure tested with no issue noted.The reported condition was not verified.Should additional relevant information become available, a supplemental report will be submitted.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRO-VOLUME EXTENSION SET
Type of Device
FILTER, INFUSION LINE
Manufacturer (Section D)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
haina san cristobal
Manufacturer (Section G)
BAXTER HEALTHCARE - DOMINICAN REPUBLIC
carretera sanchez km 18.5
parque industrial itabo, piisa
haina, san cristobal
DR  
Manufacturer Contact
25212 w. illinois route 120
round lake, IL 60073
2242702068
MDR Report Key8960076
MDR Text Key156449185
Report Number1416980-2019-04791
Device Sequence Number1
Product Code FPB
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K113227
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 10/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number2N3350
Device Lot NumberASKU
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/18/2019
Is the Reporter a Health Professional? Yes
Date Manufacturer Received09/25/2019
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage N
Patient Sequence Number1
-
-