• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZOLL MANUFACTURING CORPORATION LIFEVEST WCD 4000 SYSTEM; WEARABLE CARDIOVERTER DEFIBRILLATOR Back to Search Results
Model Number WCD 4000
Device Problems Device Alarm System (1012); Defibrillation/Stimulation Problem (1573)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 07/12/2019
Event Type  malfunction  
Manufacturer Narrative
Device evaluation of electrode belt sn (b)(4) has been completed.The reported problem (gong alarms and cannot baseline) has been confirmed.Upon investigation the electrode belt failed a therapy electrode recognition test.The cause for the failure was isolated to an open pulse wire in the cable connecting the distribution node (dn) and ecg "b".The root cause for the open wire was excessive force.No adverse event resulted from the defective electrode belt.
 
Event Description
A us distributor reported that a patient was unable to be baselined and that the patient was receiving frequent gong alarms.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LIFEVEST WCD 4000 SYSTEM
Type of Device
WEARABLE CARDIOVERTER DEFIBRILLATOR
Manufacturer (Section D)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer (Section G)
ZOLL MANUFACTURING CORPORATION
121 gamma drive
pittsburgh PA 15238 3495
Manufacturer Contact
brooke arnold
121 gamma drive
pittsburgh, PA 15238-3495
4129683333
MDR Report Key8961444
MDR Text Key156623080
Report Number3008642652-2019-06991
Device Sequence Number1
Product Code MVK
UDI-Device Identifier00855778005135
UDI-Public00855778005135
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor
Reporter Occupation Other
Type of Report Initial
Report Date 09/03/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/04/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Model NumberWCD 4000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/06/2019
Date Manufacturer Received08/09/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured05/15/2014
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-