• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ENCORE MEDICAL L.P. RSP; SOCKET, INSERT 32MM STD. RSP HUMERAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ENCORE MEDICAL L.P. RSP; SOCKET, INSERT 32MM STD. RSP HUMERAL Back to Search Results
Model Number 508-00-032
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Post Operative Wound Infection (2446)
Event Date 08/06/2019
Event Type  Injury  
Manufacturer Narrative
The reason for this revision surgery was due to an infection.The previous surgery and the surgery detailed in this investigation occurred 1.7 years apart.The healthcare professional indicated there was a significant adverse event to the patient.There was no delay in surgery and another suitable device was available for use.The revision surgery was completed as intended.The device was disposed of at the hospital and not made available to djo surgical for examination.A review of the implant device history records shows that the reported component used in the previous surgery, when released for use, met design and manufacturing requirements.There were no nonconforming material reports associated with the product that may have contributed to the reported event.The device was verified to have gone through an acceptable sterilization process and was within its expiration date at the time of the previous surgery.Customer complaint history of the reported device showed no present trends or on-going issues that are in need of review.The root cause of this complaint was a revision surgery due to an infection.There were no findings during this investigation that indicate that the reported device was the source or had a direct connection with the patient's infection.There are multiple factors that may contribute to an infection that are outside of the control of djo surgical.There are no indications of a product or process issue affecting implant safety or effectiveness.
 
Event Description
Revision surgery - due to a two year post-operation infection; related to immuno suppressant drugs.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
RSP
Type of Device
SOCKET, INSERT 32MM STD. RSP HUMERAL
Manufacturer (Section D)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer (Section G)
ENCORE MEDICAL L.P.
9800 metric blvd
austin TX 78758 5445
Manufacturer Contact
teffany hutto
9800 metric blvd
austin, TX 78758-5445
MDR Report Key8962411
MDR Text Key156781367
Report Number1644408-2019-00826
Device Sequence Number1
Product Code PHX
UDI-Device Identifier00888912024648
UDI-Public(01)00888912024648
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K051075
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 09/04/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date08/25/2022
Device Model Number508-00-032
Device Catalogue Number508-00-032
Device Lot Number855C2133
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 08/06/2019
Initial Date FDA Received09/04/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured09/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age57 YR
-
-