Model Number M00542251
Device Problem Failure to Fire (2610)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 08/15/2019
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during hemostasis, the band could not be deployed.There was no difficulty experienced when setting up the device.The procedure was completed with another speedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Manufacturer Narrative
Problem code 2610 for the reportable issue of bands failed to deploy.Investigation results: received one speedband superview super 7 for analysis and the ligator head was not returned with the device.It was noticed that the crimp was present on the trip wire and the trip wire was secured in the handle assembly slot when received.The suture was not damaged and was attached to the trip wire loop.The returned product looked in good condition without evidence of damages.A functional evaluation was performed by rotating the handle knob 180 degrees, an audible click was heard and indents were felt.No visible issue was noted with the handle assembly.As the ligator head was not returned, it was not possible to confirm what the customer alleged.Factors encountered during the procedure, such as the interaction between the device and the patient anatomy and/or the manner as the device was manipulated, could have caused the event reported by the customer.However, based on the information provided, the most probable cause of the reported issue cannot be established due to lack of evidence.Therefore, the most probable root cause for the complaint event is no problem detected since the reported device complaint or problem cannot be confirmed.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
 
Event Description
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during hemostasis, the band could not be deployed.There was no difficulty experienced when setting up the device.The procedure was completed with another speedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPEEDBAND SUPERVIEW SUPER 7
Type of Device
LIGATOR, HEMORRHOIDAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
MDR Report Key8971527
MDR Text Key156809210
Report Number3005099803-2019-04434
Device Sequence Number1
Product Code FHN
UDI-Device Identifier08714729201960
UDI-Public08714729201960
Combination Product (y/n)N
PMA/PMN Number
510K EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 10/16/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/06/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/10/2020
Device Model NumberM00542251
Device Catalogue Number4225
Device Lot Number0023465295
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/28/2019
Date Manufacturer Received09/24/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age60 YR
Patient Weight65
-
-