Model Number M00542251 |
Device Problem
Failure to Fire (2610)
|
Patient Problem
No Consequences Or Impact To Patient (2199)
|
Event Date 08/15/2019 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Although the suspect device has been received, the evaluation has not been completed.Therefore, the cause of the reported malfunction has not been determined.Upon completion of the failure analysis of the complaint device, if there is any further relevant information from that review, a supplemental mdr will be filed.
|
|
Event Description
|
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during hemostasis, the band could not be deployed.There was no difficulty experienced when setting up the device.The procedure was completed with another speedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
|
Manufacturer Narrative
|
Problem code 2610 for the reportable issue of bands failed to deploy.Investigation results: received one speedband superview super 7 for analysis and the ligator head was not returned with the device.It was noticed that the crimp was present on the trip wire and the trip wire was secured in the handle assembly slot when received.The suture was not damaged and was attached to the trip wire loop.The returned product looked in good condition without evidence of damages.A functional evaluation was performed by rotating the handle knob 180 degrees, an audible click was heard and indents were felt.No visible issue was noted with the handle assembly.As the ligator head was not returned, it was not possible to confirm what the customer alleged.Factors encountered during the procedure, such as the interaction between the device and the patient anatomy and/or the manner as the device was manipulated, could have caused the event reported by the customer.However, based on the information provided, the most probable cause of the reported issue cannot be established due to lack of evidence.Therefore, the most probable root cause for the complaint event is no problem detected since the reported device complaint or problem cannot be confirmed.A review of the device history record (dhr) was performed and confirmed that this device met all material, assembly and performance specifications.
|
|
Event Description
|
It was reported to boston scientific corporation that a speedband superview super 7 device was used in the esophagus during an esophagogastroduodenoscopy (egd) procedure performed on (b)(6) 2019.According to the complainant, during hemostasis, the band could not be deployed.There was no difficulty experienced when setting up the device.The procedure was completed with another speedband superview super 7 device.There were no patient complications reported as a result of this event.The patient condition at the conclusion of the procedure was reported to be stable.
|
|