H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.The following were reviewed as part of this investigation: patient severity, applicable previous investigation(s), sample (if available), applicable fmea documents, labeling, and applicable manufacture records.Based on a review of this information, the following was concluded: the device was returned to the service facility for evaluation.During evaluation, the reported issue of "site rite 8 isn't holding a charge.Unit turns off at ~30% battery." was confirmed.The scanner shuts off prematurely after ac power is removed.Reported issue root cause: the root cause of the reported issue is an internal li-ion battery failure.The device was serviced, tested, and returned to the customer.A history review of serial number: (b)(6) showed no other similar complaint(s) from this serial number.
|