MAUDE Adverse Event Report: INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD 2000 SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

INTEGRA LIFESCIENCES CORPORATION OH/USA MAYFIELD 2000 SKULL CLAMP; SKULL CLAMPS AND HEADREST SYSTEMS Back to Search Results
Catalog Number A2000
Device Problem Unintended Movement (3026)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
The device was not yet received by the manufacturer for evaluation.The plant investigation is in progress and a supplemental medwatch report will be submitted upon completion of the investigation.
 
Event Description
A sales representative reported on behalf of the customer that the a2000 mayfield 2000 skull clamp had a slight movement in the rocker arm when it was locked in place.The issue was found upon inspection; hence it was not used on a patient.No adverse event reported.
 
Manufacturer Narrative
Additional information - d10, g4, g7, h2, h3, h6, h10.Udi -(b)(4).The device was returned for evaluation and the unit was received with the hex bushing worn and the lock still moved after the unit was lock down and needed to be replaced.The reported complaint was confirmed.The complaint was likely caused by wear and tear over time/ improper handling.The definite root cause could not be reliably determined.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MAYFIELD 2000 SKULL CLAMP
Type of Device
SKULL CLAMPS AND HEADREST SYSTEMS
Manufacturer (Section D)
INTEGRA LIFESCIENCES CORPORATION OH/USA
4900 charlemar drive
4900 charlemar drive
cincinnati OH 45227
MDR Report Key8972473
MDR Text Key190340300
Report Number3004608878-2019-00834
Device Sequence Number1
Product Code HBL
Combination Product (y/n)N
PMA/PMN Number
K932807
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/19/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberA2000
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/19/2019
Initial Date Manufacturer Received 08/19/2019
Initial Date FDA Received09/06/2019
Supplement Dates Manufacturer Received10/08/2019
Supplement Dates FDA Received10/18/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-