• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MPRI SPECIFY SURESCAN; STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF) Back to Search Results
Model Number 977C165
Device Problem Impedance Problem (2950)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 09/05/2019
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received by a manufacturer representative (rep) regarding an implantable neurostimulator (ins) lead.It was reported that there was an impedance issue on the lead.The rep took the lead out of the battery multiple times, wiped the contacts down, re-inserted the lead and the same 2 electrodes kept showing impedance issues.The lead ports in the battery were switched and the same 1 electrode showed impedance issues.The lead was replaced and the issue was resolved.The lead was expected to be returned for analysis.No further complications were reported.
 
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPECIFY SURESCAN
Type of Device
STIMULATOR, SPINAL-CORD, IMPLANTED (PAIN RELIEF)
Manufacturer (Section D)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer (Section G)
MPRI
road 149 km 56.3
villalba PR 00766
Manufacturer Contact
lisa woodward clark
7000 central avenue ne rcw215
minneapolis, MN 55432
7635263920
MDR Report Key8977507
MDR Text Key156985493
Report Number2649622-2019-16225
Device Sequence Number1
Product Code GZB
UDI-Device Identifier00643169464599
UDI-Public00643169464599
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 11/15/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/09/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date07/19/2023
Device Model Number977C165
Device Catalogue Number977C165
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/12/2019
Is the Reporter a Health Professional? No
Date Manufacturer Received10/21/2019
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/19/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-