Catalog Number 05.001.201
Device Problem Unexpected Shutdown (4019)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 08/26/2019
Event Type  malfunction  
Manufacturer Narrative
Udi: (b)(4).Concomitant med products and therapy dates: battery handpiece device and power module devices (b)(6) 2019.Reporter¿s complete mailing address was not provided.The actual device has been returned and is currently pending evaluation.Once the device has been evaluated, a supplemental medwatch report will be sent accordingly.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
This is report 2 of 4 for the same event.It was reported from (b)(6) that the two battery handpiece devices and two power module devices did not work, stopped operation and were difficult to connect to the handpiece.It was not reported if the device was used in surgery, or if there was patient involvement.It was not reported if there were any delays in a surgical procedure, or if a spare device was available for use.It was not reported if there were any injuries, medical intervention or prolonged hospitalization.All available has been disclosed.If additional information were to become available, a supplemental medwatch will be submitted accordingly.
 
Manufacturer Narrative
Depuy synthes is submitting this report pursuant to the provisions of 21 cfr, part 803.This report may be based on information which depuy synthes has not been able to investigate or verify prior to the required reporting date.This report does not reflect a conclusion by fda, depuy synthes or its employees that the report constitutes an admission that the device, depuy synthes, or its employees caused or contributed to the potential event described in this report.The actual device was returned for evaluation.During evaluation it was observed that the device passed all manufacturing specifications.Therefore, the reported condition was not duplicated and confirmed.An assignable root cause was not determined.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BATTERY HANDPIECE/MODULAR FOR TRS
Type of Device
MOTOR, SURGICAL INSTRUMENT, AC-POWERED
Manufacturer (Section D)
DEPUY SYNTHES PRODUCTS LLC
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key8982612
MDR Text Key157170721
Report Number8030965-2019-68150
Device Sequence Number1
Product Code GEY
UDI-Device Identifier07611819977815
UDI-Public07611819977815
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/27/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2019
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.201
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/27/2019
Date Manufacturer Received10/03/2019
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-