Model Number 804018
Device Problem Patient-Device Incompatibility (2682)
Patient Problem Osteolysis (2377)
Event Date 09/27/2018
Event Type  Injury  
Manufacturer Narrative
Product is absorbed, so no samples can be provided.Risk management files for the product have been updated to include the originally unidentified risk regarding the occurrence of osteolysis in some patients implanted with the biodegradable cement restrictors.Additionally, a risk benefit analysis was completed in may 2014 that concluded the overall medical benefits of the synplug products may not outweigh individual residual risk of osteolysis, primarily due to the lack of osteolysis reports for other cement restrictor types and options available to surgeon.Based on all available data at that time, no causal relationship between the use of the products and the reported instances of osteolysis could be demonstrated.To date, there are no conclusive findings from this or similar complaint investigations, or from technical or clinical information in the literature that proves or disproves a causal relationship between synplug® & optiplug® biodegradable cement restrictors or the materials they are manufactured with, and periprosthetic osteolysis (or fractures as a result).There are also no data or findings that would suggest that only some subset of all the products manufactured might be affected.A risk/benefit analysis of two reported fractured, 63 reported osteolysis without the need for medical intervention and 100,000+ patients benefiting from a successful surgery it can be concluded that the benefit outweighs the risk.The finding of osteolysis surrounding the distal cement restrictor is unexpected, and undesirable; however, periprosthetic osteolysis in total hip arthroplasty is a well-known problem that is typically a multifactorial process and may be identified through routine radiographic follow-up.Root cause is unknown.
 
Event Description
Initial hip replacement (hip arthroplasty) surgery involving a hip prosthesis and alleged synplug cement restrictor, left side occurred in 2012.Follow-up review (7 years) determined osteolysis distal to the tip of the prosthesis.Revision surgery occurred in summer 2018 to replace the shaft on the left side.The product was absorbed and will not be returning for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Type of Device
CEMENT OBTURATOR
Manufacturer (Section D)
ISOTIS ORTHOBIOLOGICS, INC
2 goodyear
MDR Report Key8984780
MDR Text Key161922793
Report Number2090010-2019-00006
Device Sequence Number1
Product Code LZN
Combination Product (y/n)N
Reporter Country CodeSZ
PMA/PMN Number
K010840
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign
Reporter Occupation Physician
Type of Report Initial
Report Date 09/10/2019
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received09/10/2019
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Device Model Number804018
Device Lot Number112.76
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received08/28/2019
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other; Required Intervention;
Patient Age84 YR
-
-