The device history record (dhr) review confirmed that the device met all material, assembly and performance specifications.Device analysis was performed on the returned sample, and it was revealed that the stent proximal stabilizer was broken.Functional testing was performed, the distal tip of the stabilizer was cut, to remove the stabilizer from the catheter; friction was observed during the attempt to remove the stabilizer.The device was stated to be in good condition and was prepared according to the dfu specifications.The reported issue is covered in the device directions for use.The risk of the reported event is documented in the risk documentation and there are current controls to mitigate the risk of the as reported event.Based on the investigation findings, probable the proximal end of the stabilizer was kinked and broken as a result of handling of the device during the clinical procedure, therefore a cause of handling damage will be assigned to the instigation.
|